Table 2 Comparison of the conveniency and accuracy for positioning of the femoral component of the implants between PSI and CI groups

From: Computed tomography-based patient-specific cutting guides used for positioning of the femoral component of implants during unicompartmental knee arthroplasty: a cadaver study

Variables

Osteotomy time (minutes)

Coronal alignment of prostheses (o)

Sagittal alignment of prostheses (o)

PSI group (n = 25)

3.12 ± 0.65

1.43 ± 0.93

8.80 ± 0.65

CI group (n = 25)

4.33 ± 0.73

2.65 ± 1.50

6.29 ± 1.88

Statistic value (t)

-12.194

-3.666

6.959

P-value

< 0.001

0.001

< 0.001

Increased (or decreased) fold*

-1.20 ± 0.20

-1.19 ± 1.75

2.50 ± 1.81